Fluke 718 -850 mbar to 7 bar Pressure Calibrator
- RS Stock No.:
- 351-2815
- Mfr. Part No.:
- FLUKE-718 100
- Manufacturer:
- Fluke
- RS Stock No.:
- 351-2815
- Mfr. Part No.:
- FLUKE-718 100
- Manufacturer:
- Fluke
Specifications
Technical data sheets
Legislation and Compliance
Product Details
Find similar products by selecting one or more attributes.
Select all | Attribute | Value |
|---|---|---|
| Brand | Fluke | |
| Maximum Pressure Measurement | 7bar | |
| Product Type | Pressure Calibrator | |
| Battery Type | Alkaline | |
| Accuracy | ±0.015 % | |
| Minimum Pressure Measurement | -850mbar | |
| Resolution | 0.1 mbar | |
| Data Logging | No | |
| Model Number | 718 | |
| Display Type | LCD | |
| Current Loop Measurement | Yes | |
| Power Source | Battery | |
| Weight | 992g | |
| Minimum Operating Temperature | -10°C | |
| Maximum Operating Temperature | 55°C | |
| Battery Life | 20h | |
| Height | 66mm | |
| Length | 216mm | |
| Standards/Approvals | No | |
| Width | 94mm | |
| Hazardous Area Certification | NO | |
| Select all | ||
|---|---|---|
Brand Fluke | ||
Maximum Pressure Measurement 7bar | ||
Product Type Pressure Calibrator | ||
Battery Type Alkaline | ||
Accuracy ±0.015 % | ||
Minimum Pressure Measurement -850mbar | ||
Resolution 0.1 mbar | ||
Data Logging No | ||
Model Number 718 | ||
Display Type LCD | ||
Current Loop Measurement Yes | ||
Power Source Battery | ||
Weight 992g | ||
Minimum Operating Temperature -10°C | ||
Maximum Operating Temperature 55°C | ||
Battery Life 20h | ||
Height 66mm | ||
Length 216mm | ||
Standards/Approvals No | ||
Width 94mm | ||
Hazardous Area Certification NO | ||
RoHS Status: Exempt
Fluke Pressure Calibrator, ±0.015% Accuracy, 7 Bar Maximum Pressure Measurement - FLUKE-718 100
This pressure calibrator is a handheld instrument designed to provide a compact, battery-powered solution for calibrating transmitters, gauges and switches in field and workshop settings. It combines a pressure source with milliamp measurement capability and an integrated pump that resists fluid ingress, allowing technicians to generate and measure pressure while powering loop devices.
Features & Benefits
• ±0.015% mA accuracy for precise current loop verification
• Pressure range from -850 mbar to 7 bar for wide coverage
• High-resolution 0.1 mbar resolution for fine pressure discrimination
• LCD display showing simultaneous pressure and current readings
• Battery operation delivering up to 20 h typical use per set of batteries
• Integrated pump with serviceable valve assembly for field cleaning
• Pressure range from -850 mbar to 7 bar for wide coverage
• High-resolution 0.1 mbar resolution for fine pressure discrimination
• LCD display showing simultaneous pressure and current readings
• Battery operation delivering up to 20 h typical use per set of batteries
• Integrated pump with serviceable valve assembly for field cleaning
**Applications**
• Suitable for commissioning and calibrating pressure transmitters on site
• Ideal for testing and adjusting pressure switches in process systems
• Used for functional verification of gauges across -850 mbar to 7 bar
• Can be used for loop-powered sensor checks and mA calibration
• Used with non-corrosive gases for field pressure validation
• Ideal for testing and adjusting pressure switches in process systems
• Used for functional verification of gauges across -850 mbar to 7 bar
• Can be used for loop-powered sensor checks and mA calibration
• Used with non-corrosive gases for field pressure validation
What battery type does it use and how many are required?
It uses alkaline 9V cells and requires two batteries to power all functions.
How does the pump cope with accidental fluid exposure?
The pump features protective check valves and a removable valve assembly enabling in-field cleaning using isopropyl alcohol and simple tools.
What measurement capability exists for current loops?
It measures 0 mA–24 mA with 0.001 mA resolution and an accuracy of 0.015%+1 count and can drive 20 mA into 1000 Ω when battery voltage exceeds 6.8 V.
Which pressure units and ranges are supported via the built-in pump?
The instrument supports multiple units including psi, bar, mbar, kPa, inH2O and mmHg and covers negative pressures to -850 mbar up to its full-scale positive range.
What environmental limits should users observe during operation?
Operate between -10 °C and 55 °C with specified relative-humidity limits and up to 3000m altitude
storage extends beyond these values.
Supplied with
protective yellow holster, test lead set, crocodile clips, 9V alkaline battery (fitted)