TRACOPOWER THM 15WI DC-DC Converter, 15V dc/ 1A Output, 9 → 36 V dc Input, 15W, Through Hole, +85°C Max Temp

Bulk discount available

Subtotal (1 tube of 10 units)*

PHP36,495.20

(exc. VAT)

PHP40,874.62

(inc. VAT)

Add to Basket
Select or type quantity
Temporarily out of stock
  • Shipping from August 13, 2026
Need more? Click ‘Check delivery dates’ to find extra stock and lead times.
Tube(s)
Per Tube
Per Unit*
1 - 4PHP36,495.20PHP3,649.52
5 - 9PHP35,400.36PHP3,540.036
10 +PHP34,338.35PHP3,433.835

*price indicative

RS Stock No.:
146-5950
Mfr. Part No.:
THM 15-2413WI
Manufacturer:
TRACOPOWER
Find similar products by selecting one or more attributes.
Select all

Brand

TRACOPOWER

Output Voltage

15V dc

Input Voltage Range

9 → 36 V dc

Power Rating

15W

Input Voltage Nominal

24 V dc

Output Current

1A

Mounting Type

Through Hole

Isolated

Yes

Number of Outputs

1

Output Voltage Adjustment Range

-10 %, +20%

Package

1.6 x 1 in

Isolation Voltage

5kV ac

Length

40.6mm

Width

25.4mm

Depth

10.2mm

Series

THM 15WI

Load Regulation

0.002

Maximum Temperature

+85°C

Standards Met

ANSI/AAMI ES 60601-1, CB, EMC, EN 55011, EN 55032, EN 61000-4-2, EN 61000-4-3, EN 61000-4-4, EN 61000-4-5, EN 61000-4-6, EN 61000-4-8, ES 60601-1, IEC 60601-1, IEC 60601-1-2, IEC/EN 60601-1-2, MIL-STD-810F

Efficiency

88%

Medical Approved

Yes

Minimum Temperature

-40°C

COO (Country of Origin):
TW
The THM 15WI series is a range of medical 15 Watt DC/DC converters in 1.6” x 1.0” plastic package and with wide 2:1 input voltage range. They provide a reinforced isolation system for 5000 VACrms isolation and a very low leakage current of less than 2.5 μA. The units are approved to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk management file. Design and production conform to the quality management system ISO 13485.

I/O isolation 5000 VACrms rated for 250 VACrms working voltage

Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP

Risk management process according to ISO 14971 including risk management file

Extended operating temperature range –40°C to 85°C.

EMC compliance to IEC 60601-1-2 4th edition and EN55032 class A

Related links